MD40DD-P301SEQ
MAX line
Pharmacy refrigerator
Dimensions (HxWxD): 2025x600x620mm
Gross capacity: 385 L
Compliance with DIN13277
Refrigerant: R600a
Data transmision via RS485 network
Temperature recording and transfer via USB
Electronic control system
LED display
Emergency power supply to the controller
Protection against low temperature
Protection against contact with a cold surface
Name | Value |
---|---|
Pharmacy refrigerator | MD40DD-P301SEQMAXXXX1SN4B |
EAN code | 4770104040092 |
Gross capacity, l | 385 |
Net capacity, l | 270 |
Factory temperature setting, °C | +5 |
Fluctuation, °C | 0,5 |
Gradient, °C | 2,9 |
Increase in temperature from +5 °C to +10 °C (empty, +25 °C TU) | 40 min. |
Decrease in temperature from +25 °C to +5 °C (empty, +25 °C TU) | 40 min. |
Recovery time after 1 min of door opening (empty, +25 °C TU, according to EN 60068-3) | 8 min. |
Positive temperature deviation from the setpoint | 1.3 |
Negative temperature deviation from the setpoint | -1.1 |
Heat output, kJ/h | 285 |
Ambient temperature range, °C | from +10 to +35 |
Airborne acoustical noise emissions, dB(A) re1 pW | 50 |
Refrigerant | R600a |
Daily consumption, kWh/24h | 1.903 |
Annual energy consumption, kWh/year | 695 |
Rated voltage, V | 220 - 240 |
Rated frequency, Hz | 50 |
Self closing door | Yes |
Access port | Yes |
Calibration block | Yes |
Volt-free alarm contact | Yes |
Cooling system | Fan forced cooling |
Defrost | Automatic |
Interior light | LED ceiling lighting |
Malfunction warning signal | Optical and acoustical |
Power failure alarm | Immediately upon power failure for min. 12h |
Connection to the external alarm system | Yes |
Data logger | Integrated, USB interface |
Protection against low temperature | Yes |
Number of shelves | 5 adjustable +1 bottom |
Shelf size (W x D), mm | 380 x 495, bottom 445 x 180 |
Maximum load per shelf, kg | 40 |
Shelf material | Wire shelves coated by white polymer paint |
Side wall material | Steel, white painted |
Inner cabinet material | HIPS polystyrene |
Type of door | Double panel selective glass door |
Type of lock | Mechanical |
Adjustable foot at the front | 2 |
Wheels behind | 2 |
External dimensions (H x W x D), mm | 2025 x 600 x 620 |
Internal dimensions (H x W x D), mm | 1645 x 500 x 418 |
Package dimensions (H x W x D), mm | 2080 x 620 x 630 |
Weight net, kg | 80.5 |
Weight gross, kg | 86 |
Pharmacy refrigerator packed on a pallet | MD40DD-P301SEQMAXXXX1SN4A |
EAN code | 4770104040092 |
Dimensions (H x W x D), mm | 2210 x 620 x 710 |
Weight gross, kg | 94,5 |
CERTIFICATIONS, APPROVALS: DIN 13277, IEC 60068-3
Description
SNAIGĖ pharmaceutical refrigerators ensure the maximum stable internal temperature and its minimum fluctuations. Such conditions are suitable for the storage of many drugs and medicines.
MAX line refrigerators have all the essential features and performance functions required to meet the performance criteria of the DIN 13277 standard. This standard provides the requirements and test specifications for medical refrigerators
SNAIGĖ medical refrigerators are tested according to EN 60068-3 in order to ensure the stability and constancy of the internal temperature. The test method involves measuring the temperature at 15 different points inside the refrigerator.
Optical and acoustic alarms will warn if the temperature rises or the power supply stops for any reason. In a power failure, the controller will operate for 12 hours. Temperatures are recorded and transmitted via USB. We also offer XWEB monitoring system for a data transfer via RS485 network.
SNAIGĖ pharmaceutical refrigerators are manufactured using only non-harmful, non-ozone-depleting and non-greenhouse-effect refrigerating and insulating materials. Natural resource-saving materials and electricity from renewable energy sources are used in the refrigerator manufacturing process
STANDART FEATURES
Temperature stability
The refrigerator is designed to maintain maximum stable temperature and minimal fluctuations. The maximum possible temperature change within the entire internal space of the refrigerator is only ± 2.9 °C. Thus, the internal microclimate is maintained, creating suitable conditions for the storage of medications and medicines.
Integrated data logger
The memory of the MD40DD-P301SEQ electronic controller stores the last 100 alarm records with the date and duration of the alarm, as well as the temperature measurement summary with the measurement date and time. These saved data can be transferred to a USB storage device.
Warning of electricity supply failure
In the event of the electricity supply failing, the battery of the MD40DD-P301SEQ provides sufficient energy for the registration of temperature readings and the operation of the warning signal (audio and visual) for the next 12 hours.
Warning temperature alarm
In order to ensure the maximum safety of stored medicines and medications, the integrated warning temperature alarm sends a warning signal with sound and image when a certain temperature value exceeds or falls below the set limit.
Warning register
Warning signals must be monitored to ensure the stored medicines or medications retain their quality and integrity. Important data are recorded in the register: date of warning, duration, highest temperature. The memory of the MD40DD-P301SEQ electronic controller stores the last 100 warning records.
+2 °C temperature limit
If the temperature drops below +2 °C for some critical reasons, the electronic controller immediately activates additional temperature maintenance functions along with a warning signal in order to protect the temperature-sensitive medicines and medications stored in the refrigerator.
LAN interface
SNAIGĖ medical refrigerators can be connected to a LAN network (using external systems) through the integrated LAN interface RS485. It is convenient to monitor, control and digitally document the temperature of medicines stored in the refrigerator.
Access to an additional external sensor
An additional external sensor can be attached if required for additional control of the temperature of medication storage in the refrigerator. SNAIGĖ medical refrigerators are equipped with a 12.5 mm diameter access for an additional external sensor, which helps avoid any negative effect of the cables of the external sensor on the sealing strip of the door, the tightness of the refrigerator and the risk to the personnel caused by the cables of the sensor.
Self-closing door
The door of the SNAIGĖ medical refrigerator closes automatically, without any risk of the door bouncing. This ensures convenience and safety for medical personnel in their daily use of the refrigerator, and also completely eliminates the risk of the refrigerator door being accidentally left open. The self-closing door contributes to maintaining a stable temperature, as well as supporting the integrity of stored medicines and medications.
Calibration block
Calibration is necessary so that both the internal air temperature measured and recorded by the electronic controller and the temperature of the medical products stored in the refrigerator match each other as closely as possible.
Protection against contact with cold surfaces
The evaporator of the refrigerator has a special cover so that medications and medicines do not have direct contact with the evaporator and do not get cold. In addition, this protection against contact with cold surfaces increases the temperature stability.
Mechanical door lock for a safe storage of medicines
Changeable direction of the door opening
Replaceable door sealing strip
To be ordered separately
XWEB monitoring system for MD40DD-P***SEQ
- Can monitor up to 100 units of control panels connected to one network ;
- XWEB internal data storage capacity ensures the information storage for up to 1 year;
- If XWEB is connected to LAN it allows to monitor information from any device connected to the internet;
- Adjustments to controller parameters can be made remotely;
- Alarm indication can be external and send signals via LAN (email, SMS or FAX) to the responsible person;
- Report generation in pdf format;
- Recorded data transfer from the device to the USB.
To be ordered separately
- Branding
- XWEB monitoring system
FUNCTIONS AND DESCRIPTION OF THE CONTROL PANEL:
- Temperature recording and alarm recording functions;
- Transfer of saved data to a USB storage (it goes together with refrigerator);
- Emergency 12- hour power supply and alarm function;
- Setting the maximum allowable temperature limit;
- Setting the minimum allowable temperature;
- Possibility to connect an external alarm system;
- Possibility to connect to external temperature and other parameters monitoring system XWEB (purchased separately).